<?xml version="1.0" encoding="UTF-8"?>



<records>

  <record>
    <language>eng</language>
          <publisher>Oriental Scientific Publishing Company</publisher>
        <journalTitle>Biosciences Biotechnology Research Asia</journalTitle>
          <issn>0973-1245</issn>
            <publicationDate>2016-06-15</publicationDate>
    
        <volume>12</volume>
        <issue>Spl,Edn.1</issue>

 
    <startPage>25</startPage>
    <endPage>36</endPage>

	 
      <doi>10.13005/bbra/1601</doi>
        <publisherRecordId>12550</publisherRecordId>
    <documentType>article</documentType>
    <title language="eng">The Effects of Backscattered Radiation into Beam Monitor Chamber: Study of 6 and 18 MV Conventional and Removed Flattening Filter Clinical Accelerator</title>

    <authors>
	 


      <author>
       <name>Mansour Zabihzadeh</name>

 
		
	<affiliationId>1,2</affiliationId>
      </author>
    

	 


      <author>
       <name>Seyyed Rabi Mahdavi</name>


		
	<affiliationId>3</affiliationId>

      </author>
    

	 


      <author>
       <name>Mohammad Reza Ay</name>

		
	<affiliationId>4</affiliationId>
      </author>
    

	 


      <author>
       <name>Zahra Shakarami</name>

		
	<affiliationId>1</affiliationId>
      </author>
    


	


	
    </authors>
    
	    <affiliationsList>
	    
		
		<affiliationName affiliationId="1">Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.</affiliationName>
    

		
		<affiliationName affiliationId="2">Department of Radiotherapy and Radiation Oncology, Golestan Hospital, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.</affiliationName>
    
		
		<affiliationName affiliationId="3">Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran.</affiliationName>
    
		
		<affiliationName affiliationId="4">Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.</affiliationName>
    
		
		
	  </affiliationsList>






    <abstract language="eng">In some linear accelerators (Linac), the collected charges in beam monitor
chamber (BMC) is partly caused by the backscattered particles from the accelerator
components downstream the BMC that influence the Linac output factors. In the intensity
modulated radiation therapy technique, the desired dose distribution can be achieved
through an unflattened beam. Although removing the flattening filter provides some
advantages, the amount of backscatter radiation into BMC can be changed. In this study,
contribution of backscattered particles into the BMC response of a Varian 2300 C/D Linac
with and without a flattening filter was determined for 6, 18 MV photon beams. The
experimental procedure included telescopic method and calculation procedure consisted
of Monte Carlo simulation (MCNPX, version 2.4.0), were used to investigate the
contribution of backscattered particles into the BMC performance. Our results showed a
2.3 % and 3 % increase in backscatter for a 0.5 × 0.5 cm2 field compared to a 40 × 40 cm2
field for 6 MV and 18 MV, respectively. The energy deposition from backscattered radiation
is mainly caused by backscattered electrons. Removing the flattening filter did not change
the BMC performance for a conventional Linac with a flattening filter. However, this
result was not valid for small fields (e.g. 0.5 × 0.5 cm2, 18 MV). The corrected backscatter
factors is necessary to taking into account the contribution of backscattered radiation in
the monitor chamber response for small fields in the case of the free flattening filter
Linacs (18 MV).</abstract>

    <fullTextUrl format="html">https://www.biotech-asia.org/vol12_nospl_edn1/the-effects-of-backscattered-radiation-into-beam-monitor-chamber-study-of-6-and-18-mv-conventional-and-removed-flattening-filter-clinical-accelerator/</fullTextUrl>



      <keywords language="eng">
        <keyword>Backscatter radiation; Flattening filter; Linac; MC simulation; Telescopic analysis</keyword>
      </keywords>

  </record>
</records>